Important Safety Notice VISTA 120 CMS SpO2 Alarm : Software Patch for Vista CMS required

 

 

Affected Medical Device : Vista 120 and Central Monitoring Station (CMS) using CMS SW1.64

 

MDA Reference Number : MDA/FCA/2021-P130

 

MDA Registration Number : GC3840819-37203

 

Decription : Draeger has become aware of an event during which an exceeding of SpO2 alarm thresholds triggered an alarm at the bedside monitor Vista 120 as expected but that this did not lead to visual or audible alarming at the Central Monitoring Station (CMS). Numeric values were displayed correctly.

 