The list below contains Medical Device’s Field Corrective Action for the month of February 2026. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

  • The return of the medical device to the establishment;
  • Modification of the medical device;
  • Exchange of the medical device;
  • Destruction of medical device; or
  • Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

        MEDICAL DEVICE FIELD CORRECTIVE ACTION LISTING FEBRUARY 2026

No.

Date Received

Title of FCA

Affected Medical Device

MDA Reference Number

MDA Registration Number

Local Establishment Contact Detail

1

02/02/2026

Xpert® Carba-R Sample Reagent Vial with Incorrect Volume

XPERT CARBA-R ASSAY

 MDA/FCA/P1640-40868003-2026

IVDC6457221-74902

RADIOMETER MALAYSIA SDN. BHD.

ahila.padmanathan@radiometer.my

2

03/02/2026

(PHC26-01.A.OUS) Protein S Ac kit lot 03628 – Protein S Ac Reagent lot 562954 may be cloudy after reconstitution.

SIEMENS PROTEIN S AC

 MDA/FCA/P1641-25745633-2026

IVDC4691624-166628

SYSMEX (MALAYSIA) SDN. BHD.

 cheah.kityin@sysmex.com.my

3

04/02/2026

Discontinuation use and repair of TU2000 TRYTABLE

MMT TRYTABLE TU2000

 MDA/FCA/P1646-99461683-2026

GA2807724-178157

MEDI-LIFE (M) SDN. BHD.

christine@medi-life.com.my

4

05/02/2026

FCA2525: TS-10: Lift Drop Incident during Rack Jam Clearence

XR-SERIES AUTOMATED HEMATOLOGY ANALYZER

 MDA/FCA/P1647-89136998-2026

IVDB6506724-170755

SYSMEX (MALAYSIA) SDN. BHD.

cheah.kityin@sysmex.com.my

5

05/02/2026

Olympus Emphasis on Strict Adherence to Instructions for Use (IFU)

EUS ULTRASONIC VIDEOSCOPE SYSTEM

 MDA/FCA/P1648-63811962-2026

GB79514818718

OLYMPUS (MALAYSIA) SDN. BHD.

hideki.nagai@olympus.com

6

05/02/2026

FA-001491- Beckman Coulter has determined that some DxC 700 AU and AU5800 Clinical Chemistry Analyzers may unexpectedly generate some errors due to Microsoft® SQL Server® 2022 Express issue. If this issue occurs the system may stop analyzing, and cause delays in reporting patient test results to physicians.

DxC 700 Analyzer

 MDA/FCA/P1649-92028478-2026

IVDA5467424-161450

BECKMAN COULTER MALAYSIA SDN. BHD.

song01@beckman.com

7

05/02/2026

LIFECODES LSA Class I Stability Failure

LIFECODES SINGLE ANTIGEN KITS

 MDA/FCA/P1650-78301862-2026

IVDC75845124518

BMS DIAGNOSTICS (M) SDN BHD

rqa.my@biomedglobal.com

8

06/02/2026

Field Safety Notice - STA Liatest D-Di Plus. Stago has identified a positive bias in D-Dimer results across the analytical range for the affected lots.

STA® - LIATEST® D-DI PLUS

MDA/FCA/P1651-19481110-2026

IVDC5347522-90795

ALL EIGHTS (M) SDN BHD

all8@alleights.com.my

9

06/02/2026

015 FSCA IMC 26-03 IMMULITE 2000 EPO and IMMULITE EPO Control Module | EPO Adjustor and Control Module Open Vial Instability

IMMULITE 2000 EPO

MDA/FCA/P1653-60601548-2026

IVDB1287363017

SIEMENS HEALTHCARE SDN. BHD.

aili.siow@siemens-healthineers.com

10

09/02/2026

CVG-21-Q3-21 Rev AC, HVAD Pump Failure to Restart Algorithm C

HEARTWARE™ VENTRICULAR ASSIST DEVICE (HVAD™) ACCESSORIES SYSTEM

MDA/FCA/P1654-58219598-2026

GD2555223-122718

MEDTRONIC MALAYSIA SDN BHD cherie.ang@medtronic.com

11

10/02/2026

Olympus to Provide Impacted Customer With a Copy of the Instructions for Use (IFU)

CELONLAB ENT SYSTEM

MDA/FCA/P1659-81299459-2026

GC35106930618

OLYMPUS (MALAYSIA) SDN. BHD.

hideki.nagai@olympus.com

12

11/02/2026

Important Security Notice for Centricity Universal Viewer Centricity Universal Viewer (GB5328025-213680)

CENTRICITY UNIVERSAL VIEWER

MDA/FCA/P1663-28353720-2026

GB5328025-213680

GE HEALTHCARE SDN. BHD

premarket.my@gehealthcare.com

13

12/02/2026

GEM PREMIER 5000 BLOOD GAS SYSTEM Issue: Performance issue with specific GEM PAKs for the GEM Premier 5000 with iQM2 (Intelligent Quality Management) may experience an increased incidence of “Process Control Solution Not Detected” (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring a new GEM PAK insertion.

GEM PREMIER 5000 BLOOD GAS SYSTEM

MDA/FCA/P1664-90928076-2026

IVDC33500201118

STRAITS SCIENTIFIC (M) SDN BHD

straits@3s.com.my

14

13/02/2026

Device Name: HEMOSIL REAGENTS - HemosIL Factor V Leiden Issue: Reduction in the on-board instrument stability claim for all current and future lots of HemosIL Factor V Leiden when used on the ACL TOP Family, ACL TOP Family 50 Series, and ACL TOP Family 70 Series hemostasis systems.

HEMOSIL REAGENTS

 MDA/FCA/P1665-66157857-2026

IVDB75996368519

STRAITS SCIENTIFIC (M) SDN BHD

straits@3s.com.my

15

14/02/2026

FA1514 MiniMed Pump Placement with Respect to Infusion Site

MINIMED 700 INSULIN PUMP AND KIT

 MDA/FCA/P1666-83333705-2026

GC8922821-64034

MEDTRONIC MALAYSIA SDN BHD

 cherie.ang@medtronic.com

16

14/02/2026

FA1514 MiniMed Pump Placement with Respect to Infusion Site

MINIMED 780G SYSTEM AND ACCESSORIES

 MDA/FCA/P1667-59518771-2026

GC4190021-62126

MEDTRONIC MALAYSIA SDN BHD

cherie.ang@medtronic.com

17

14/02/2026

FA1514 MiniMed Pump Placement with Respect to Infusion Site

PARADIGM VEO INSULIN PUMP

 MDA/FCA/P1668-71818633-2026

GC8904122-88631

MEDTRONIC MALAYSIA SDN BHD

cherie.ang@medtronic.com

18

14/02/2026

                FA1514 MiniMed Pump Placement with Respect to Infusion Site

MINIMED 640G SYSTEM AND ACCESSORIES

 MDA/FCA/P1669-52527916-2026

GC8430922-106837

MEDTRONIC MALAYSIA SDN BHD

cherie.ang@medtronic.com

19

14/02/2026

FA1514 MiniMed Pump Placement with Respect to Infusion Site

THE MINIMED PARADIGM INSULIN PUMP

 MDA/FCA/P1670-21003774-2026

GC8737122-88537

MEDTRONIC MALAYSIA SDN BHD

cherie.ang@medtronic.com

20

19/02/2026

2026-CC-SRC-002: Trilogy Evo, Trilogy Evo O2, Trilogy EV300 Non-pneumatic Nebulizer Use, Flow Sensor Nebulized Aerosol Deposition, and Obstruction Alarm

TRILOGY EVO

 MDA/FCA/P1673-56947816-2026

GC2627520-39891

PHILIPS MALAYSIA SENDIRIAN BERHAD

shireenn.tiam@philips.com

21

19/02/2026

2026-CC-SRC-002: Trilogy Evo, Trilogy Evo O2, Trilogy EV300 Non-pneumatic Nebulizer Use, Flow Sensor Nebulized Aerosol Deposition, and Obstruction Alarm

TRILOGY EV300

 MDA/FCA/P1675-32908560-2026

GC5498920-50666

PHILIPS MALAYSIA SENDIRIAN BERHAD

shireenn.tiam@philips.com

22

20/02/2026

Field Safety Notification for Locking adapter/s to the Leksell Gamma Knife

LEKSELL GAMMA KNIFE (LGK)

 MDA/FCA/P1680-43676500-2026

GC9257822-110526

NYPRAX BUSINESS SOLUTIONS

kenna@nyprax.com

25

23/02/2026

Confirming Intended Reagent is Loaded During Manual Load Procedures

VITROS IMMUNODIAGNOSTIC PRODUCTS FERTILITY & PREGNANCY HORMONES & PROTEINS

 MDA/FCA/P1681-84081900-2026

IVDB7258323-129195

T T MEDICAL MANAGEMENT SDN BHD

ar@ttmedical.com.my

26

23/02/2026

VITROS™ Immunodiagnostic Products Progesterone 2 Calibrators – Potential for Calibration or Quality Control Failures

VITROS IMMUNODIAGNOSTIC PRODUCTS PROGESTERONE 2

 MDA/FCA/P1682-14057287-2026

IVDB3167525-214895

T T MEDICAL MANAGEMENT SDN BHD

ar@ttmedical.com.my

29

25/02/2026

ABL90 FLEX PLUS - Analyzer may abort capillary samples

ABL90 FLEX

MDA/FCA/P1687-55374265-2026

IVDC5590952317

RADIOMETER MALAYSIA SDN. BHD

ahila.padmanathan@radiometer.my.

30

26/02/2026

Device Name: HemosIL Factor XIII Antigen Issue: Reduction in the on-board instrument stability claim for all in-date and future lots of HemosIL Factor XIII Antigen (Part No. 0020201300) when used on the ACL TOP Family, ACL TOP Family 50 Series, and ACL TOP Family 70 Series hemostasis systems.

HEMOSIL FACTOR XIII ANTIGEN

 MDA/FCA/P1688-60639833-2026

IVDC2356823-153931

STRAITS SCIENTIFIC (M) SDN BHD

straits@3s.com.my

31

27/02/2026


017 FSCA ACHC 26-04 Atellica CH Enzymatic Hemoglobin A1c (A1c_E) | Potential for Depressed Results with Atellica CH Enzymatic Hemoglobin A1c when Processing Atellica CH Revised C‑Reactive Protein on the Same Analyzer

ATELLICA® CH ENZYMATIC HEMOGLOBIN A1C (A1C_E)

MDA/FCA/P1689-24384977-2026

IVDB9845619-35149

SIEMENS HEALTHCARE SDN. BHD.

aili.siow@siemens-healthineers.com